Cryogenic Elaton
Program Directive
To design, fabricate, and test a spaceflight qualified tunable Fabry-Perot
etalon.
Program Description
Michigan Aerospace Corporation is presently contracted with NASA Langley
Spaceflight Center to design, fabricate and test a spaceflight qualified
tunable Fabry-Perot etalon. This device will be operated at 77 Kelvin
from 7 - 11 microns. Tuning is accomplished via this use of ferroelectric
single crystals and will provide a range of travel of approximately
20 microns at operating temperature. Zinc Selenide is used for all transmitting
elements. Titanium (Ti-6Al-5V Grade 5) is used for the etalon mount
due to its low density, high strength, and coefficient of thermal expansion.
|
Cryogenic
Etalon Requirements and Parameters
|
|
Parameter
|
Desired
|
Minimum
|
Design Projected
|
|
Wavelength
|
9-12um
|
-
|
-
|
|
Gap
|
-
|
-
|
3.7mm
|
|
Finesse
|
20
|
12
|
19.3
|
|
Transmission
|
11%
|
8%
|
8.10%
|
|
Resolution
|
0.07 cm-1
|
6.32 Angstroms at 9.5um
|
6.32 Angstroms at 9.5um
|
|
Reflectivity
|
0.8
|
0.6-0.9
|
0.85
|
|
Dynamic Range
|
25um
|
5um
(1/2 wavelength)
|
20.5
|
|
Free Spectral Range
|
Maximize
|
-
|
12.2nm
|
|
|